| Inflow mechanism | Fluid flows from an elevated reservoir under gravity; hydrostatic head provides the driving pressure. | There is no pump actively sensing and regulating intra-articular pressure. |
| Pressure source | Pressure depends primarily on the vertical distance between the fluid surface and the joint, as well as fluid-line resistance. | Changing reservoir height changes the available hydrostatic pressure, but does not provide automatic pressure compensation. |
| Illustrative static head | For water-like fluid, 50 cm, 75 cm, and 100 cm of vertical head correspond to approximately 37, 55, and 74 mmHg, respectively. | These are theoretical static values; actual pressure at the joint can be lower during flow because of tubing, cannula, and outflow resistance. |
| Pressure adjustment | Typically adjusted by raising or lowering the reservoir and managing inflow or outflow pathways. | Adjustment is relatively simple, but changes are manual and may not respond immediately to changing surgical conditions. |
| Flow behavior | Flow varies with hydrostatic head, tubing configuration, cannula size, fluid viscosity, and resistance at the outflow path. | Flow and joint distension may fluctuate when instruments are inserted, outflow changes, or the reservoir level falls. |
| Equipment setup | Generally uses an elevated fluid bag or reservoir, tubing, and compatible inflow connections; it does not require a powered pressure-control unit. | Setup is comparatively straightforward, but reservoir positioning and secure suspension are important. |
| Potential advantages | Simple operating principle, no powered pressure-regulation mechanism, and straightforward visual monitoring of the fluid supply. | May suit procedures where basic gravity-driven irrigation is appropriate and the clinical team can manage inflow and outflow manually. |
| Limitations | No active pressure feedback; performance can vary as reservoir height, fluid level, and pathway resistance change. | Requires ongoing attention to joint distension and fluid egress. It may be less suitable when stable, actively regulated pressure is required. |
| Safety considerations | Pressure at the joint is not determined by reservoir height alone; anatomy, leakage, outflow, and procedure-specific factors also matter. | Use the device according to its instructions and the surgical team’s protocol; monitor the patient and joint throughout the procedure. |